To adjust the lateral vertical light projector, proceed as follows:
For Orthoralix 9200/Plus
1.
Make sure that the system is in the reset position.
2.
Insert the light phantom (a) (code 4519 190 00922) into the
bite block holder (b).
3.
Switch on the light and verify that lateral vertical light is
phantom
Right
Wrong
Gendex Dental Systems 4519-190-71193-7
Wrong
F-63
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
ORTHORALIX 9200 Service Manual
4.
If not, remove the cover (d) of the light projector by un-
screwing three screws (c), accessible from the rear side of
the panoramic cassette carriage.
5.
Access to the lateral vertical light projector and:
of the light beam
screw (e) and rotate the light projector (f).
6.
Reinstall the cover (c) of the light projector.