F.5.4
Replacing optoswitches
Two types of optoswitches could be mounted in the system:
Type 1
looking at the optoswitch from the front it is possible to see two
letters printed on it: an E and a S.
For the replacement, the wires must be connected as shown in
the figure below:
Rear view
White
Gendex Dental Systems 4519-190-71193-7
E
Green
S
Brown
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
Type 2
Looking at the optoswitch from the front it is possible to see that
one part is grey and the other one is black.
For the replacement, the wires must be connected as shown in
the figure below:
White
No other special precautions are required for replacement of the
optoswitches, as they are mounted at fixed positions.
After replacement of the X, Y, or R movements optoswitches pro-
ceed with the relevant cam adjustment, verifying the centering of
the system using the image quality phantom (see Section C and
Section F of this Service Manual).
If a collimator optoswitch is defective, it is necessary to replace
the optoswitch board PA2: verify then that the three positions
of the collimator (panoramic, transcan and ceph) are correctly
selected by the optoswitches.
F-38
Rear view
Grey
Green
Brown