Gendex ORTHORALIX 9200 Service Manual page 113

Panoramic x-ray
Hide thumbs Also See for ORTHORALIX 9200:
Table of Contents

Advertisement

ORTHORALIX 9200 Service Manual
F.5.6
Replacing the tubehead
For the replacement of the tubehead proceed as follows:
1.
Remove the front cover (a) of the tubehead by unscrewing
two screws (b) and by disengaging the spring latches from
the rear cover (c).
2.
Remove the rear cover (c) by unscrewing the four Allen
screws (d).
3.
Disconnect the wires from the terminal block 600 of the
tubehead.
4.
Disconnect the ground cable from the tubehead.
5.
Disconnect connector X50 from the collimator.
Gendex Dental Systems 4519-190-71193-7
F-44
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
6.
Remove the collimator assembly from the tubehead, by
unscrewing the four screws (g).
7.
Remove the tubehead (e) by unscrewing four hexagonal
head screws (f).
8.
Proceed backwards for remounting.
(03.0)E