ORTHORALIX 9200 Service Manual
3. Ground AC service to L-bracket using an electrician supplied
tting listed for grounding, or other approved method in
accordance with the local electrical code (see Figure 2). A 0.875"
hole is proved in the bracket suitable for ½" conduit.
Figure 2: 'L' Bracket
4. Route Line, Neutral, and Earth Ground Conductors under side
cover bracket and under front column cover bracket (see Figure
3).
Figure 3: Routing under brackets
Gendex Dental Systems 4519-190-71193-7
B-30
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
5. Terminate the conductors in the terminal strip and strain
relief using the cable clamp located below the terminal strip
(see Figure 1 for location of terminal strip and cable clamp).
6. Attach the front column cover assuring the wiring routes
through the opening provided in the bottom left corner (see
Figure 4).
Figure 4: Front cover in place; arrow shows where wiring
will route.
7. Attach the side cover to the column.
(06.0)E