F.6.4
Adjusting the light projectors
the eyes directly to the laser beam.
1.
Make sure that the system is in the reset position.
2.
Insert the light phantom (a) (code 4519 190 00922) into the
bite block holder (b).
3.
the phantom.
Gendex Dental Systems 4519-190-71193-7
F-61
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
ORTHORALIX 9200 Service Manual
4.
(c) by unscrewing two Allen screws (d).
5.
Disconnect the connector X31 from the PC1 control board.
6.
screw (f): then rotate the light projector to correct the hori-
7.
is adjusted (if not, repeat the procedure from point (4) to (6).
(03.0)E