Gendex ORTHORALIX 9200 Service Manual page 43

Panoramic x-ray
Hide thumbs Also See for ORTHORALIX 9200:
Table of Contents

Advertisement

ORTHORALIX 9200 Service Manual
6.
Fix the cable (n) on the column by using the cable retiners
(o) and (p) supplied.
7.
Feed the spiral cable in the Z-carriage through the hole (q)
provided.
Gendex Dental Systems 4519-190-71193-7
B-26
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
8.
Connect the cable to the connector X33 of the PC1 board:
X33-1
RED WIRE
X33-2
BLACK WIRE
X33-3
WHITE WIRE
X33-4
YELLOW WIRE
9.
Connect the cable to the terminal block of the column, as
shown in the figure below.
10.
Connect the external lights to the terminal block of the col-
umn (for more information see also the principle diagram
(Z-15).
11.
Mount back the right lateral colum cover taking care to
feed the external light cable connection through the hole
provided (see figure below).
12.
Mount back the following covers:
PC1 system control board).
Yellow
Hole
White
Black
Red
X-Ray
Ready
(03.0)E