E.7.1
Verify the anodic current
mA
NOMINAL VALUE (V)
3
1.57
5
7
8
10
11
12
13
7.33
15
E.7.2
Verify the kv
kV
NOMINAL VALUE (V)
70
7.0
72
7.2
78
7.8
80
8.0
82
8.2
E.7.3
Verify the dose rate
kV
mA
10
70
10
80
10
MEASURED VALUE (V)
MEASURED VALUE (V)
DOSE RATE
s
Gy/s
R/m
1
1
1
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
SELECTED TIME ( s )
1.0
2.50
12 (Standard Panoramic)
E.7.5
Verify the reproducibility
EXPOSURE N°
R / MIN
1
2
3
5
7
8
10
RADIATION LEAKAGE
E.7.7
Verify the earth resistance
EARTH RESISTENCE
E.7.8
Verify the leakage current
S1
S2
S3
OFF
OFF
ON
OFF
ON
ON
ON
OFF
ON
ON
ON
ON
OFF
OFF
OFF
OFF
ON
OFF
ON
OFF
OFF
ON
ON
OFF
during exposure
E-5
Service Manual
MEASURED TIME ( s )
MEASURED DOSE
Gy / s
MAX 250 µGy/h (28mR/h) A 1m
MAX 0,1 Ohm
EARTH LEAKAGE CURR. (mA)
MAX.
MEASURE
< 5
< 5
< 5
< 5
< 10
< 10
< 10
< 10
< 10