1.
Make sure that the system is in the reset position.
2.
Insert the light phantom (a) (code 4519 190 00922) into the
bite block holder (b).
3.
Switch on the light and verify that lateral vertical light is
phantom
Right
Wrong
Gendex Dental Systems 4519-190-71193-7
Wrong
F-64
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
4.
If not, remove the cover (a) by removing the three screws (b)
and then pull it out to disengage from the spring latches.
5.
Access to the lateral vertical light projector and:
Allen screw (e) and rotate the light projector (f)
the Allen screw (c) and move the whole light projector
assembly (d).
6.
Reinstall the cover (c).