Service Manual
C.5.2.2
Orthoralix 9200 DDE
To check the alignment of the X-ray field in transcan mode pro-
ceed as follows:
1.
2.
Select the program X-ray alignment and press Dummy Run;
a static exposure.
3.
Select the transcan position on the collimator and on the
operator panel.
4.
5.
Switch on the PC where the Orthoralix 9200 DDE is connected
to.
6.
four options will be displayed:
7.
Gendex Dental Systems 4519-190-71193-7
C-6
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
8.
Select Static image by cliking on it.
9.
appear on the monitor:
the X-ray field respect to the digital sensor.
paragraph Adjusting the X-ray collimator of the Section F
10.
on Static image.
11.
12.
-
Stop and then
Stop and then on OK.