F.5
Replacement procedures
F.5.1
Replacing multipole cable
For the replacement of the multipole cable follow the procedure
below:
Before starting the replacement of the multipole
cable, be sure that no dangerous voltages are
present.
1.
Remove the top longer cover (a) of the rotating arm by
unscrewing the six screws and the top short cover (b) by
unscrewing the four screws.
2.
Remove the top cover (e) of the Z-carriage by unscrewing
the six Allen screws.
3.
Remove the lower cover (c) of the Z-carriage by unscrewing
the four screws.
4.
Remove the central cover (d).
Gendex Dental Systems 4519-190-71193-7
F-33
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER
ORTHORALIX 9200 Service Manual
5.
Remove the front panel with mirror (f) by unscrewing the
two screws (g) and disconnect the centering lights connector
X31 from the PC1 control board.
6.
Remove the headrest cover (h) by unscrewing the four screws
(i).
7.
Disconnect the connector X13 from the headrest.
8.
Remove the headrest assembly (l) by unscrewing the three
Allen screws (m).
(03.0)E