Ametek Thermox 2000 ATEX Series User Manual page 186

Table of Contents

Advertisement

Location
(Hex/Decimal) Variable Name
61/97
A-24 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Description
IQ
Which oxygen alarm has been set and
Alarm_flags
why
bit 0 – high oxygen, process alarm
bit 1 – low oxygen, process alarm
bit 2 – high combustible, process
alarm
bit 3 – high methane, process alarm
bit 4 – high oxygen, alarm #2
bit 5 – low oxygen, alarm #2
bit 6 – high combustible, alarm #2
bit 7 – high methane, alarm #2
bit 8 – high oxygen, alarm #3
bit 9 – low oxygen, alarm #3
bit 10 – high combustible, alarm #3
bit 11 – high methane, alarm #3
bit 12 – Future use
bit 13 - Out of Range, for SA/F sensor
bit 14 - Future use
bit 15 - Future use
S2
Alarm_flags
Which oxygen alarm has been set and
why
bit 0 – low oxygen, alarm #3
bit 1 – high oxygen, alarm #3
bit 2 – low oxygen, alarm #4
bit 3 – high oxygen, alarm #4
bit 4 – high combustion, alarm #5
bit 5 – high combustion, alarm #6
bit 6 – high methane, alarm #7
bit 7 – high methane, alarm #8
bit 8 – Future use
bit 9 – Future use
bit 10 – Future use
bit 11 – Future use
bit 12 – Future use
bit 13 - Future use
bit 14 - Future use
bit 15 - Future use
Access
Read only
Read only

Advertisement

Table of Contents
loading

This manual is also suitable for:

Thermox wdg-hpiiThermox wdg-hpiic