Set Cal Timers - Set Auto Timers
This menu option allows you to set the frequencies for automatic calibra-
tions or verifications, including the time when the first calibration or
verification should begin. You have the option of setting the cycle in
either hours or days.
If you schedule a verification and calibration at the same time, the
verification will be performed first, then the calibration.
NOTE
Select Set Auto Timers from the Set Cal Timers menu.
Select Auto Cal Cycle if defining an automatic calibration cycle.
Select Auto Verify Cycle if defining an automatic verification cycle.
Select Set Cycle in DAYS or Set Cycle in HOURS, depending on
whether you want to set the frequency of calibrations or verifications
in days or hours.
If you are setting the auto timers for the first time, the Series 2000
software will automatically prompt you to enter the current time
NOTE
and date.
You will then be prompted to enter the frequency for the calibration or
verification cycle (either in days or hours, depending on what you
selected in Step 3 above):
Cal Cycle = 01 days
New Value = xx
Enter the number of days (or hours) for the automatic calibration or
verify cycle. As you enter the values, they overwrite the XXs on the
display. To enter 1 day or 1 hour, enter 01; otherwise, the system will
assume you are entering 10 days (or hours).
You will then be prompted to enter the time of day when the first
calibration of the cycle should begin:
Cal time is YY:YY
Enter new time Hr: Mn