Location
(Hex/Decimal) Variable Name
Iout_flags
Description
S2
Analog outputs flags (1=true, 0=false)
bit 0 – analog output 3 tracks during
calibration
0 is hold analog output 3 during
calibration
bit 1 – analog output 3 tracks during
verify
0 is hold analog output 3 during
verify.
bit 2 – analog output 3 set for 0 to
20 ma, 0 is 4 to 20 ma
bit 3 – Future use
bit 4 – analog output 4 tracks during
calibration
0 is hold analog output
calibration.
bit 5 – analog output 4 tracks during
verify
0 is hold analog output 4 during
verify.
bit 6 – analog output 4 set for 0 to
20 ma, 0 is 4 to 20 ma
bit 7 – Future use
bit 8 – analog output 1 tracks during
calibration
0 is hold analog output 1 during
calibration
bit 9 – analog output 1 tracks during
verify
0 is hold analog output during verify.
bit 10 – Analog output 1 is set to 0-
20, 0= Output 1 is set to
bit 11 – Future use
bit 12 – analog output 2 tracks during
calibration
0 is hold analog output 2 during
calibration
bit 13 – analog output 2 tracks during
verify
0 is hold analog output during verify.
bit 14 – analog output 2 set for 0 to
20 ma, 0 is 4 to 20 ma.
bit 15 – Future use
Access
R/W
4 during
4-20ma
Serial Communications
|
A-13