Location
(Hex/Decimal) Variable Name
03/03
Iout_flags
A-12 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Description
IQ
Analog outputs flags (1=true, 0=false)
bit 0 – analog output 1 tracks during
calibration
0 is hold analog output 1 during
calibration
bit 1 – analog output 1 tracks during
verify
0 is hold analog output during verify.
bit 2 – Future use (used on S2000)
bit 3 – Future use
bit 4 – analog output 2 tracks during
calibration
0 is hold analog output 2 during
calibration
bit 5 – analog output 2 tracks during
verify
0 is hold analog output during verify.
bit 6 – analog output 2 set for 0 to
20 ma, 0 is 4 to 20 ma.
bit 7 – Future use
bit 8 – analog output 3 tracks during
calibration
0 is hold analog output 3 during
calibration
bit 9 – analog output 3 tracks during
verify
0 is hold analog output 3 during
verify.
bit 10 – analog output 3 set for 0 to
20 ma, 0 is 4 to 20 ma
bit 11 – Future use
bit 12 – analog output 4 tracks during
calibration
0 is hold analog output
calibration.
bit 13 – analog output 4 tracks during
verify
0 is hold analog output 4 during
verify.
bit 14 – analog output 4 set for 0 to
20 ma, 0 is 4 to 20 ma
bit 15 – Future use