Ametek Thermox 2000 ATEX Series User Manual page 187

Table of Contents

Advertisement

Location
(Hex/Decimal) Variable Name
62/98
Cal_variable
63/99
Option_flags
64/100
Tspan_mv
65/101
Tzero_mv
66/102
Mthvalue
67/103
M_store
68/104
Cal_temp
69/105
CJTemp
6A/106
Box_temp
6B/107
Cmb_Kfact
6C/108
Mth_Kfact
6D/109
Description
The calculated percent value that 0 mv
would correspond to with the cell that
is currently being used.
can vary between different cells.
is calculated during calibration and
used in the oxygen calculations.
Option flags (1=true, 0=false)
bit 0 – combustible detector installed
bit 1 – Data module installed
Dual current output
S2
installed
bit 2 – 2 span value RCU
auto cal option installed
S2
bit 3 – Methane detector installed
3-in-1 option installed
bit 4 – Dig – ma module installed
bit 5 – HART module installed
bit 6 – RS485 module installed
bit 7 – 3 span value RCU
Invalid option module
Cell mv value measured with the span
gas flowing during last calibration.
Cell mv value measured with the zero
gas flowing during calibration.
Last calib temperature
Cold Junction temperature
Box Temperature
Combustible K Factor
IQ
Methane K Factor
Access
Read only
This value
It
Read only
IQ
IQ
IQ
S2
IQ
IQ
IQ
IQ
S2
Read only
R/W
Read only
Read only
R/W
Read only
Read only
R/W
R/W
Serial Communications
|
A-25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Thermox wdg-hpiiThermox wdg-hpiic

Table of Contents