Figure 4-4.
Sensor temperatures
by analyzer type.
4-16 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Sensor Config.
Use this menu to select the sensor you will be using with the Series 2000.
When selecting a sensor, use the table below to see the temperature each
type of sensor uses. If your type of sensor is not listed, find the operating
temperature for your sensor, then choose the sensor below that uses that
operating temperature.
Use caution when selecting the type of sensor used with the Series
!
2000 Control Unit. Selecting the wrong sensor will damage sensor
CAUTION
components and could invalidate your warranty coverage.
Configurating the sensor
1.
Select Sensor Config. from the Setup menu.
2.
Select the sensor you will use with the Series 2000 control unit (see
the list of options in the table above). If your sensor is not listed, call
the factory for the correct operating temperature for your sensor
type.
Sensor
WDG-Insitu Probe
Normal Cell Range
High Cell mV Range
WDG-IV or HP Series
Normal Cell Range
High Cell mV Range
CEM/O2
Normal Cell Range
High Cell mV Range
Operating Temperature
(degrees Celsius)
615
650
695
824
695
812