4-24 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Type a number between 1 and 100, then press the Enter key. Note
that you should enter 01 to enter 1, 07 to enter 7, 70 for 70, etc.
The default value is 100 (no filtering).
NOTE
Select Function
The Select Function menu option allows you to select what type of
information the analog output port should track. Choices are as follows:
% oxygen
Cell temperature
Thermocouple millivolts (T/C mv)
Cell millivolts
Combustibles (optional)
The combustibles option is available only for current outputs 3 and 4,
and cannot be set for analog output ports 1 or 2.
Selecting the function
Choose Select Function from the Analog Range menu after selecting
an analog output port.
Select the function you want the selected analog output port to track
(see bullets in Description section above).
The Series 2000 control unit then prompts you with the Set Current
Range menu option where you choose the range of readings that the 0-
to-20 mA or 4-to-20 mA analog output represents (see the "Set Current
Range" section in this chapter for details).