(Hex/Decimal) Variable Name
02/02
Config_flags
A-10 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Description
IQ
System configuration flags (1=true,
0=false)
bit 0 – relays energize on alarm.
not, they deenergize.
Bit 1 – auto calibration time is set,
an auto calibration will happen at the
set time if the auto calibration is
enabled
bit 2 – auto calibration timer is
enabled, since this is enabled, an
auto calibration will occur at the
auto calibration time.
bit 3 – real time clock is set.
bit 4 – Step Function enabled (only
for RSS)
bit 5 – auto verify time is set, an
auto verify will happen at the set
time if the auto verify is enabled.
bit 6 – auto verify timer is enabled,
if this is enabled an auto verify will
occur at the auto verify time.
bit 7 – auto calibration cycle is
programmed in hours, 0 if in days
bit 8 – auto verify cycle time is
programmed in hours, 0 if in days
bit 9 – Future use
bit 10 – Future use (used on S2000)
bit 11 – Future use (used on S2000)
bit 12 – Future use (used on S2000)
bit 13 – Future use (used on S2000)
bit 14 – WDG type sensor, 0 is HP type
sensor
bit 15 – combustibles are measure in
percent,0 = combustibles are measured
in PPM
If
(used on S2000)
Access
R/W