Figure 5-2.
Set Cal Timers menu
options.
Set Cal Timers
Set Cal Timers
The Set Cal Timers submenu from the Calibrate key allows you to do the
following:
Set auto timer for desired auto calibration/verification cycle times (Set
•
Auto Timers).
Set the current time and date (Set Time and Date).
•
Enable or disable the auto calibration cycle (Auto Cal On/Off).
•
Define how long cal gases should flow through the sensor during
•
automatic calibrations or verifications (Cal Gas Duration).
Define a recovery duration, during which time the sensor switches
•
from measuring calibration gas readings to measuring process read-
ings without affecting the analog output or triggering of alarms
(Recovery Duration).
Set Auto Timers
Set Time & Date
Auto Cal On/Off
Cal Gas Duration
Recovery
Duration
Calibration | 5-21