Out # 1 will HOLD
during Verify
Set Current Mode
The Set Current Mode menu option allows you to define either a 0-to-20
or 4-to-20 mA analog output range for the selected analog output port.
This output is dependent on both the range of readings you define (see
the "Set Current Range" section) and on the function you track on the
analog output (see the "Select Function" section).
1.
Select the Set Current Mode menu option from the Analog Range
key after selecting an analog output port.
2.
Select 4-to-20 mA Mode or 0-to-20 mA Mode.
The following will then appear on the display:
Out # 1 is set for
Out # 1 is set for
Default Selection: 4-to-20 ma
Output Filtering Option
The Output Filtering menu option allows you to filter out quick transi-
tions in readings to allow a smoother analog output.
1.
Select Output Filtering from the Analog Range key after selecting an
analog output port.
You will be prompted to enter an output filtering value:
[100 =FAST 1 = SLOW]
Fltr = XX New #?
where XX is the current filter value. Note that 100 equals no filtering,
and 1 equals the highest possible filter value.
or
4 - 20 ma
or
0 - 20 ma
Controller / User Interface | 4-23