3-4 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX

Mounting the Sensor

You must attach the probe, exhaust tube, and probe filter assembly to
the sensor before mounting the sensor to the process.
NOTE
The ambient temperature range for the sensor is -20 to 60
sor mounting dimensions are provided in Figure 3-2.
The three methods of mounting the WDG-HPII to the process are as fol-
lows:
Directly weld the Thermox supplied mounting plate to the process. In
1.
this case, the pipe nipple attached to the mounting plate is inserted
directly into the process. This is the preferred method of installation.
Weld the Thermox-supplied mounting plate with attached pipe nipple
2.
to a pipe nipple that you provide.
Mount the sensor using a flange that you provide. For this type of
3.
mounting, you don't need the Thermox supplied mounting plate.
o
C. Hinged sen-