Ametek Thermox 2000 ATEX Series User Manual page 181

Table of Contents

Advertisement

Location
(Hex/Decimal) Variable Name
29/41
Recov_time
2A/42
O2_gas1
2B/43
O2_gas2
2C/44
Co_gas1
2D/45
Co_percent_gas
2E/46
Me_percent_gas
2F/47
Cal_cgas1
30/48
Cal_span
31/49
Ver_cgas1
32/50
Ver_span
33/51
O2_span
34/52
Cal_cgas2
35/53
Cal_zero
36/54
Ver_cgas2
Description
Amount of time for recovery after a
calibration in seconds
Amount of time for recovery after a
calibration in MMSS
Oxygen span gas setpoint value, always
in percent
Oxygen zero gas setpoint value, always
in percent
Combustibles PPM span gas setpoint
S2
value
Combustibles span gas setpoint
Combustibles percent span gas setpoint
S2
value
Methane percent span gas setpoint
value
Oxygen span gas setpoint for last
calibration
Oxygen span gas reading
calibration
Oxygen span gas setpoint for last
verify
Oxygen span gas reading from last
verify
Oxygen span gas millivolts during last
calibration
Oxygen zero gas setpoint value for
last calibration
Oxygen zero gas reading from last
calibration
Oxygen setpoint value used during last
verify
IQ
S2
IQ
for last
Serial Communications
Access
R/W
R/W
R/W
R/W
R/W
R/W
Read only
Read only
Read only
Read only
Read only
Read only
Read only
Read only
|
A-19