Location
(Hex/Decimal) Variable Name
Description
Access
S2
System configuration flags (1=true,
Config_flags
R/W
0=false)
bit 0 – auto verify cycle time is
programmed in hours, 0 if in days
bit 1 – Auto cal option is installed
bit 2 – Combust module is installed
bit 3 – Future use
bit 4 – Future use
bit 5 – RTD type combustible detector,
0= Hot wire detector
bit 6 – WDG type sensor, 0 is HP type
sensor
bit 7 – combustibles are measure in
percent,0 = combustibles are measured
in PPM
bit 8 – relays energize on alarm.
If
not, they deenergize.
bit 9 – auto calibration time is set,
an auto calibration will happen at the
set time if the auto calibration is
enabled
bit 10 – auto calibration timer is
enabled, since this is enabled, an
auto calibration will occur at the
auto calibration time. (0=true,
1=false)
bit 11 – real time clock is set.
bit 12 – Step Function enabled (only
for RSS)
bit 13 – auto verify time is set, an
auto verify will happen at the set
time if the auto verify is enabled.
bit 14 – auto verify timer is enabled,
if this is enabled an auto verify will
occur at the auto verify time.
(0=true, 1=false)
bit 15 – auto calibration cycle is
programmed in hours, 0 if in days
|
Serial Communications
A-11