8.5.2 Cleaning

Clean the product regularly depending on the degree of soiling and frequency of use, at least 1x per month:

8.5.2.1 Cleaning by hand

Clean plastic parts, frame parts and the chassis and wheels with a mild cleaner and a damp cloth. Dry thor­
oughly afterwards.
Dirty springs should be cleaned with a soft, dry brush to prevent wear.
Do not spray the product with a pressure washer.
Do not use any aggressive cleaners, solvents or hard brushes etc.

8.5.2.2 Cleaning in a washing chamber

1) Wash the outdoor mobility base at a maximum temperature of 60 °C for no longer than 10 minutes.
2) If necessary: remove all components in which water has accumulated.
3) Allow the mobility base to dry. Lay the mobility base onto both rear wheels so the water can drain.
INFORMATION: we recommend drying with compressed air.
4) If necessary: reinstall all components that were removed.
5) Before reusing or storing the outdoor mobility base, check that it is dry. Wipe wet components with a cloth.
6) Ensure that all labels (nameplate and warnings) are still legible. Have any labels that are not legible replaced.
Important information on cleaning
Do not use any aggressive cleaners, solvents or hard brushes etc.

8.5.3 Disinfection

Thoroughly clean the product before disinfecting.
Wipe all parts of the product with a disinfectant.
Only use colourless water-based disinfectants. Observe the instructions for use provided by the manufacturer.

9 Maintenance and repair

9.1 Maintenance

WARNING
Insufficient maintenance
Severe user injuries, damage to the product due to failure to observe maintenance intervals
► Only carry out the maintenance tasks described in this section. All other maintenance and service tasks may
only be carried out by qualified personnel.
► The functionality and operating safety of the product must be verified and a service performed at least
once per year.
► For users with a changing anatomy (for example body dimensions, weight) or users with a changing clinical
picture, have the product inspected, adjusted and serviced at least once every six months.
The function of the product should be checked before each use.
The product may not be used if defects are noted. This applies in particular in case of instability of the product
or altered driving characteristics as well as problems with the user's seating position or the stability of the seat.
Inform the qualified personnel promptly for the rectification of defects.