Charger Indicator Test; Operating The Pain Manager Infusion Pump - CME BodyGuard 595 Color Vision Operator's Manual

Pain manager infusion system
Table of Contents

Advertisement

Press
3.
confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
4.
Start infusion as i nstructed in the "Select protocol" section.
Wait untill pump displays "low battery" message
5.
6.
Press
7.
Connect the pump to a charger.
8.
Verify the red icon of battery is lit (on the charger).
Verify the red battery turns green after a minimum of 2 hours.
9.
Warning: do not let the pump operate when battery is fully discharged. Pump
may turn off during operation on fully discharged battery without alarm.

Charger Indicator Test

1.
Click the pump into the charger and connect the Charger to mains. Disconnect
the power cord from the AC power outlet.
Verify that the CHARGE indicator is OFF.
2.
3.
Connect the power cord to the AC power outlet. Verify that the CHARGE
indicator is ON.

Operating the Pain Manager Infusion Pump

NOTE: before using the BodyGuard 595 Color Vision™ for the first time please
ensure you have carried out the 'pump operation tests' as detailed in the
previous chapter. These tests should be carried out periodically to ensure all
systems are functioning correctly.
Before performing any of the set-up & period tests or to operate, program or configure
the pump you will need to access the pump using the security codes. As the BodyGuard
595 Color Vision™ is designed for pain management therapies the pump has several
BodyGuard 595™Color Vision Operator Manual
until the Self-Test screen appears. If air sensor is off, press
to turn the pump off.
PCA
10:00
Low Battery
Connect to Charger
4 – Operation
to
- 10%
25