7.
Depending upon the set flow rate and pressure
threshold, the Down Occlusion alarm will be
activated.
Press
Door Open – Alarm Test
1.
Insert a MicroSet
heard.
NOTE: for instructions on how to load and prime an administration set refer to
sections "Loading the Administration Set" & "Priming the Line".
Press
until the Self-Test screen appears. If air sensor is off, press
2.
confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
3.
Start infusion as instructed in the "Select protocol" section.
4.
Open the pump door.
5.
Door Open alarm will be activated at once.
6.
Close the door and press
Battery Test
1.
Insert a MicroSet
is heard.
NOTE: for instructions on how to load and prime an administration set refer to
sections "Loading the Administration Set" & "Priming the Line".
2.
Take the pump out of the charger.
BodyGuard 595™Color Vision Operator Manual
to mute the alarm.
TM
in the pump and close the door. Press the door until a click is
.
TM
in the pump and close the door. Press the door until a click
REGIONAL
10:00 -
Down
Occlusion
Check Set / Access
Press
key to mute
PCA
10:00
Door Open
Close door
4 – Operation
█
■
70%
to
█
■
-
70%
24