Terms Used In Manual; Intended Use; Warnings - CME Color Vision BodyGuard 595 Directions For Use Manual

Pain manager infusion system
Table of Contents

Advertisement

Terms Used In Manual

:
Warning
circumstances that could result in injury or death to the patient or operator.
Read and understand this manual and all warnings completely before operating
the BodyGuard 595 Color Vision™ Pain Manager.
indicates that the information is a caution. Cautions advise you of
:
Caution
circumstances that could result in damage to the device.
Read and understand this manual and all cautions completely before operating
the BodyGuard 595 Color Vision™ Pain Manager.
NOTE: indicates that the information that follows is additional important
information or a tip that will help you operating the BodyGuard 595 Color
Vision™ Pain Manager.

Intended Use

The BodyGuard 595 Color Vision™ Pain Manager is designed for infusion of medications
or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates
through
clinically
subcutaneous, percutaneous, intra-arterial, epidural, intrathecal, in close proximity to
nerves and into an intraoperative site (soft tissue/body cavity/surgical wound site). The
system is intended for patients who require maintenance medications, analgesics, PCA
therapy in hospital and home care environments.
The BodyGuard 595 Color Vision™ Pain Manager Infusion Pump is designed for patients
who require epidural, Regional or PCA pain management. The BodyGuard 595 Color
Vision™ Infusion System is for use by, or under the supervision of trained medical staff
in the hospital and the home care environments. *DOUBLE CHECK*

Warnings

To avoid possible personal injury or loss of life, observe the following:
Read the entire DFU before using the pump, since the text includes important
precautions.
The maximum volume that may be infused under SINGLE FAULT CONDITION is
0.1 ml.
Voltage present on internal components may cause severe shock or death upon
contact. Disconnect the Charger from the mains, prior to opening the casing. Only
trained service personnel should open the pump cover.
Make sure the pump is attached securely to the Charger, which is connected
firmly to an IV pole.
A kinked or occluded administration line may impair the operation of the pump and
the accuracy of the infusion. Before operation, verify that the administration line is
not kinked or occluded.
The BodyGuard 595 Color Vision™ should be operated only with the MicroSet
administration lines. Use of administration lines other than the MicroSet
impair the operation of the pump and the accuracy of an infusion.
BodyGuard 595 Color Vision™ Directions For Use
indicates that the information is a warning. Warnings advise you of
acceptable
routes
of
administration,
2 – Symbols, Warnings, and Cautions
including
intravenous,
TM
TM
may
16