Aux And E-Aux Analog (0-1Vdc, 0-5Vdc And 4/20Ma ) - YSI MultiDO 5400 User Manual

Monitor, control, & alarm instruments
Table of Contents

Advertisement

3.
Configure system menu. For Aux - page 102. For E-Aux - page 108.
4.
Enter Aux 1-6/E-Aux 1-8 in Sensor Setup menu
Menus → System → Sensor Setup → Aux 1-6
▲ ▼
5.
Use
to scroll and highlight submenus.
Press
6.
to select.
7.
Enable and configure submenus. See Aux/E-Aux Digital relay for digital
relay functionality - page 164. Use AquaManager to configure sensor
setup menus - page 150.
8.
Optional - Configure digital 4/20mA output - page 178.
9.
Press
to save configuration and return to Run Screen.
10. Test system output device(s) at→System→Service→Output Test - page
144 or create an event to verify configured relay(s) energize.
Notes:
-
It is important to keep the 5X00 at the Run Screen during normal opera-
tion. Alarm functionality (email alarms, activation of enabled alarm relays,
alarm icons, and buzzer) is suspended when 5X00 is not at Run Screen.
-
At sensor log Aux (1:1.00 ) is reporting open condition. Aux (1: 0.00) is re-
porting closed condition.
-
Do not enable aux digital alarms when the sensor is in the control/alarm con-
dition. Enabling aux digital alarms when sensor is in control/alarm condition
can result in aux system not properly processing existing alarm condition.
-
One relay can be assigned to multiple 5X00 systems. It is essential that the relay
configuration be verified in order to avoid conflict of operation of wired relay
devices. If not configured properly, assigning one relay to multiple systems
can result in invalid operation of 5X00.
-
Event logs are created when an aux relay energizes and de-energizes regardless
if alarms are enabled.
-
Aux alarms must be enabled to trigger General Alarm.
-
General Alarm may still be active even though aux system is no longer in alarm.
See aux relay information below and General Alarm - page 131.
-
E-Aux system datalogs are not accessible at the 5X00. They must be viewed
using AquaManager.
Aux and E-Aux Analog (0-1VDC, 0-5VDC and 4/20mA
Menus → Sensor Setup→Aux 1-6
→E-Aux 1-8
Configure analog control and alarm low and high values, control and alarm relays
and hold off time in the Sensor Setup Menu.
Configuring the 5X00
→ E-Aux 1-8
)
YSI 5X00
166
E-Aux systems use an IOEM channel as the analog input source. See IOEM/REM
E-AUX
Safety and Installation guide for wiring and input channel configuration.
Configure analog Aux and E-Aux systems
Before configuring Aux and E-Aux analog systems 1) See - "Use AquaManager to
configure system inputs and outputs" - page "Use AquaManager to configure System
inputs and outputs" on page 150. AND 2) See Control modes [Set point Control
Mode and Alarms - page 153 (Aux and E-Aux Analog set point control mode can
operate in normal or timed mode - page 157) and PID control mode - page 159.]
E-AUX
1.
Wire digital input at IOEM for E-Aux system - see IOEM/REM Safety
and Installation guide>Chapter 3>Step 5>Wire non temperature IOEM
inputs;
AND/OR
1.
Wire Aux analog (non temperature) device - page 56.
2.
Wire relay output device(s) to local instrument, or any networked 5X00
instrument - page 54 for local instrument.
AND/OR
2.
Wire relay output device(s) to REM - see IOEM/REM Safety and In-
stallation guide>Chapter 3>Step 4>Wire REM Relay Outputs. Wire any
output devices to be used for system.
3.
Configure system menu. For Aux - page 102. For E-Aux - page 108.
4.
Enter Aux 1-6/E-Aux 1-8 in Sensor Setup menu
Menus → System → Sensor Setup → Aux 1-6
Configuring the 5X00
→ E-Aux 1-8
YSI 5X00
167