Submenu Screen Types

The 5X00 includes numeric, alpha/numeric, and pull-down submenus - Figure 4.4.
See Front panel keys - page 73 , to navigate, enter, configure, and view submenus.
Figure 4.4
This symbol indicates that alpha/numeric or numeric keypad entry is required.
This symbol is used in the menu map appendix starting on 225 and in directions
for menu configuration (this chapter). Use
and
to select entries.
The remainder of this chapter provides information on 5X00 menus and submenus
and their function. Menu structure and information is provided in hierarchical order
starting with the
menu. Information on
menus is provided after the
Configuring the 5X00
▲ ▼
front panel keys to scroll
,
, and
menu sections.
YSI 5X00
80

Main Menu

Press
at the Run Screen to enter the Main Menu.
All system configuration is done through this menu. The four Main Menu submenus
are Calibration, System, Sensor Setup, and Timers.

Calibration Menu

DO sensors(s) must be user calibrated to ensure accurate measurements. Calibrate
DO sensors at the Calibration Menu. Entering the Calibration Menu de-energizes
control and alarm relays for the DO system being calibrated. Energized DO control
relays for systems not being calibrated remain energized including any/all tempera-
ture systems regardless if they are the configured temperature (temp) source for
the DO sensor being calibrated (5400 only). If necessary, disable any temperature
relays for the associated DO sensor in the Sensor Setup menu. There is a 30 second
hold off period for control and alarm relays after pressing
the calibration menu. The hold off period allows sensors time to stabilize before
processing readings for controlling and alarming purposes.
Notes:
-