Configure Aux; Probe Name; Input Type - YSI MultiDO 5400 User Manual

Monitor, control, & alarm instruments
Table of Contents

Advertisement

Configure AUX

Menus → System → Configure Aux 1-6
The 5X00 has six auxiliary (Aux) inputs. Aux inputs 1, 2, 3, 4, 5, and 6 can be con-
figured for digital or analog ( 0-1V,DC 0-5VDC, or 4-20 mA) inputs. In addition,
Aux 3, 4, 5, and 6 can be configured with a temperature sensor. Configure suffix,
minimum (Min), maximum (Max), Input Scale Min, Input Scale Max and display
format menus for non temperature analog inputs. To configure set points, control,
alarms and relays see specific sensor set up menus. See Aux digital sensor set up -
page 165. See Aux analog (non temperature) sensor set up - page 166. See Aux
and ODO temperature sensor set up - page 168. Configure Aux menus vary based
on the configured Input Type.
5400 Configure Aux 5 menu
(no Temp Source menu)

Probe Name

Menus → System → Configure Aux →Aux 1-6→Probe Name
Use the alpha/numeric keypad to name aux inputs. Configured name(s) appear
throughout menus and at Run Screen - figure 4.7- page 106.
Note:
-
Aux names can also be configured at Menus → System → Name Devices/
Sensors.
Configuring the 5X00
5500 Configure Aux 5 menu
(
Temp Source menu will be available when Tempera-
ture Input Type configured
YSI 5X00
102

Input Type

Menus → System → Configure Aux →Aux 1-6→Input Type
Select the input type for the device(s) wired to aux system(s). Aux Sensor Setup
menu selections are based on the aux input type configured at system>Configure
Aux menu. All input types are discussed in the following sections.
Digital
Menus → System → Configure Aux →Aux 1-6→Input Type→Digital
Use digital input when wiring an external device to control or alarm when the
contacts of an external switch open or close. A common application would be to
detect high or low water levels using a float switch.
Notes:
-
Suffix, Min, Max, and Display Format are not available for digital inputs.
-
In the 5X00 sensor log, aux digital "closed" is logged as Aux 0.00. Aux digital
"open" is logged as Aux 1.00.
Directions to configure Digital device at Configure Aux menu:
1.
Wire aux digital devices - page 57.
2.
Enter Aux menu
Menus → System → Configure Aux →Aux 1-6
Use
and
3.
to scroll and highlight submenus.
Press
4.
to select.
5.
Configure digital input type and name.
6.
Press
to save configuration and return to Run Screen.
Configuring the 5X00
→Digital
→0-1v
→0-5v
→4/20ma
→Aux 3-6→Input Type
→Temperature
YSI 5X00
103

Advertisement

Table of Contents
loading

This manual is also suitable for:

Multido 5500d-01Multido 5500d-02Multido 5500d-04

Table of Contents