CONTENTS SECTION 1 GENERAL DESCRIPTION......................3 1.1 K ............................3 EYPAD AYOUT SECTION 2 PREPARING THE YSI 550 ......................4 2.1 U ............................... 4 NPACKING 2.2 W ............................. 4 ARRANTY 2.3 B ..............................4 ATTERIES 2.4 T ........................ 5 URNING THE NSTRUMENT 2.5 C...
The YSI 550 DO Instrument is designed for field use and is available with cable lengths of 12 or 25 feet. The body of the probe has been manufactured with stainless steel to add rugged durability and sinking weight.
YSI 550 DO Instrument should ever need repair. 2.3 BATTERIES There are a few things you must do to prepare your YSI 550 DO instrument for use. First, locate the 4 C-size alkaline batteries that were included with your instrument. See Figure 2.
The light will power off automatically after two minutes of non-use. 2.5 CHANGING THE TEMPERATURE UNITS The YSI 550 DO Instrument can display the temperature units in either Fahrenheit or Celsius. To change the units that are displayed, turn the instrument on. Pressing both the down arrow key and the mode key at the same time will switch the units between C or F.
The YSI 550 DO Instrument’s storage chamber can be conveniently used from either side of the instrument. To attach the rubber stopper to the calibration chamber, first remove the chamber from the instrument by unscrewing the two screws. After the chamber is free from the instrument, the rubber stopper can be removed from the chamber by pulling the “tail”...
3.3 PROBE PREPARATION The YSI 550 DO probe is shipped dry. Before using the YSI 550, the protective membrane cap on the probe tip must be removed, the new membrane cap must be filled with electrolyte solution and installed.
CAUTION: Do not touch the membrane surface. 3.5 PROBE OPERATION AND PRECAUTIONS NOTE: The YSI 550DO Instrument should not be used in a purpose other than that specified by YSI Incorporated. See Section 7, Warranty and Repair for details. It is important to recognize that oxygen dissolved in the sample is consumed during probe operation.
O-ring. Turn the instrument on and allow the system to stabilize for at least 30 minutes. If, after several hours, you are still unable to calibrate, return the YSI 550 DO Instrument to a YSI authorized service center for service, See Section 7, Warranty and Repair for more information.
Along with the YSI 559 dissolved oxygen module, a YSI 5906 membrane kit, the set screw that holds the sensor in place, YSI supplies a hex key wrench to help install the probe, and an instruction sheet. See Figure 6.
Page 13
Wrench Figure 6 7. The YSI 559 DO module is shipped dry. Before using the YSI 550 with a new YSI 559 sensor, the protective membrane on the probe tip must be removed, the membrane cap must be filled with electrolyte solution and a new membrane must be installed. Follow the instructions below to install the electrolyte solution and membrane.
Before you calibrate the YSI 550 DO Instrument complete the procedures discussed in Section 2, Preparing the YSI 550 and Section 3, Preparing the Probe. You can calibrate the YSI 550 DO Instrument in either mg/L or % saturation. In the section below, you will find instructions on how to calibrate and change between these modes.
Page 15
Make sure that the DO reading (large display) is stable, then press the ENTER button. The YSI 550 DO instrument should now display CAL in the lower left of the display, the calibration value should be displayed in the lower right of the display and the current DO reading (before calibration) should be on the main display.
Page 16
% air saturation mode by pressing the MODE key. If you are working in a dark area and have difficulty reading the LCD, press and hold the LIGHT key to activate the back-light of the YSI 550. The ON/OFF key turns the instrument on or off. For best results: Each time the YSI 550 DO Instrument is turned off, re-calibrate before taking measurements.
DO errors caused by temperature compensation for measurements at ±10°C from calibration temperature: ±1% (0.08 mg/L at 25°C) • DO errors caused by temperature measurement errors: A maximum ±0.2°C temperature error is equal to ±0.5% (0.04mg/L at 25°C). YSI Incorporated YSI 550 DO Handheld Instrument...
Page 18
(r.m.s.) calculation: ½ r.m.s. error = ±[1a + 1b + 2a + 2b + 3a + 3b mg/L NOTE: This sample calculation is for a near extreme set of conditions. YSI Incorporated YSI 550 DO Handheld Instrument...
20 mg/L or below 0 mg/L salinity compensation (4). B. System requires service B. Replace membrane and Electrolyte (3.4) C. Clean anode (3) D. Clean cathode (3) E. Return system for service (7) YSI Incorporated YSI 550 DO Handheld Instrument...
Page 20
A. Sample temperature is more than A. Reduce the sample temperature C (122 B. Recalibrate using correct altitude and B. System requires service salinity compensation (4). Main display reads "Er 7" C. Return system for service (7) YSI Incorporated YSI 550 DO Handheld Instrument...
SECTION 7 WARRANTY AND REPAIR The YSI 550 DO Instrument is warranted for three years from date of purchase by the end user against defects in materials and workmanship. YSI 550 DO probes and cables are warranted for one year from date of purchase by the end user against defects in material and workmanship. Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determines to be covered by this warranty.
Page 22
AUTHORIZED U.S. SERVICE CENTERS North Region YSI Incorporated • Repair Center • 1725 Brannum Lane • Yellow Springs, Ohio • 45387 • Phone: (800) 765-4974 • (937) 767-7241• E-Mail: info@ysi.com South Region C.C. Lynch & Associates • 212 E. 2nd Street • Suite 203 • Pass Christian, Mississippi •...
Page 23
INTERNATIONAL SERVICE CENTERS YSI Incorporated • Repair Center • 1725 Brannum Lane • Yellow Springs, Ohio • 45387 • Phone: (937) 767-7241• E-Mail: info@ysi.com YSI Europe • Lynchford House • Lynchford Lane • Farnborough • Hampshire • GU146LT • Phone: (44-1252) 514711 • Fax: (44-1252) 511855 • Tlx: 858210 YSI Japan •...
We will contact the sender for instructions as to the disposition of the equipment. Disposition costs will be the responsibility of the sender. When service is required, either at the user's facility or at YSI, the following steps must be taken to insure the safety of our service personnel.
Page 25
Model No. of Device ______ Lot Number _________ Contaminant (if known) _____________________ Cleaning Agent(s) used _____________________ Radioactive Decontamination Certified? (Answer only if there has been radioactive exposure) ___ Yes ___ No Cleaning Certified By _________________________ Name Date YSI Incorporated YSI 550 DO Handheld Instrument...
Temperature Compensation: Automatic 4 alkaline C-cells Power: Size: 4.7 in. width; 9 in. length (11.9 cm x 22.9 cm) Weight with Batteries: 2 lb. (0.91 kg) Battery life: ° Over 2000 hours at 25 YSI Incorporated YSI 550 DO Handheld Instrument...
YSI Instrument away from the receiver • plug the YSI Instrument into a different outlet so that the computer and receiver are on different branch circuits. If necessary, the user should consult the dealer or an experienced radio/television technician for additional suggestions.
APPENDIX C ACCESSORIES/REPLACEMENT PARTS The following parts and accessories are available from YSI or any YSI Authorized Dealer. YSI Model Number Description Replacement DO Module 5906 6 Cap Membranes and Electrolyte Solution 5238 Probe Reconditioning Kit (sanding tool and disks for cathode cleaning)
Chlorinity = Measure of chloride content, by mass, of water. ) = 1.806550 x Chlorinity ( This table is provided for your information only. It is NOT required when calibrating the YSI 550 DO Instrument in accordance with the instructions outlined in the Section 4, Calibration.