Notes:
-
Values displayed at the run screen are based on display format resolution
configuration. Example - if the display format is 0.0 and the min value input
is 1.88, then 1.9 will be displayed at the run screen.
-
5X00 aux analog (non temp) sensor log records 0.00 resolution regardless of
the selected resolution in the aux analog display format configuration.
Configure E-Aux analog - Configure E-Aux system menu:
1.
Wire E-Aux analog device - see IOEM Safety and Installation
Guide>Chapter 3 Installation and Wiring>IOEM input configuration>Wire
analog input.
2.
Refer to and follow directions at IOEM Safety and Installation
Guide>Chapter 4>Quick Start - IOEM Configuration to confirm correct
network connectivity;
3.
Follow directions in the Quick Start section to configure IOEM channel
mode to 0-1v, 0-5v or 4/20mA.
4.
Enter E-Aux menu
Menus → System → Configure E-Aux →E-Aux 1-8
Use
and
5.
to scroll and highlight submenus to configure.
6.
Press
to select.
7.
Configure Analog system name - page 109, instrument - page 110,
channel - page 111, input type - page 111, suffix - page 113, min/
max - page 114, scale min/scale max -page 114and display format -
page 115menus. E-Aux submenus configuration is described in the
above sections.
8.
Press
to save configuration.
9.
Go to Sensor Setup Menus → Sensor Setup E-Aux 1-8 - page 166 to
enable and configure set points, control, alarm, relays, and hold off time.
10. Press
to save configuration.
Note:
-
The alarm low value must be ≥ the min value. The alarm high value must be
≤ the max value.
Configuring the 5X00
YSI 5X00
116
Temperature
Menus → System → Configure E-Aux →E-Aux 3-6→Input Type→Temperature
Configure E-Aux temperature Configure E-Aux system menu:
1.
Wire temperature device - see IOEM Safety and Installation Guide>Chapter
3 Installation and Wiring>IOEM input configuration>Wire temperature
input.
2.
Refer to and follow directions at IOEM Safety and Installation
Guide>Chapter 4>Quick Start - IOEM Configuration to confirm correct
network connectivity;
3.
Follow directions in the Quick Start section to configure IOEM channel
mode to temperature.
4.
Enter E-Aux menu
Menus → System → Configure E-Aux →E-Aux 1-8
Use
and
5.
to scroll and highlight submenus to configure.
Press
6.
to select.
7.
Configure name- see page 109 instrument - page 110, channel - page
111, input type - page 111 and display format (Fo or Co).
8.
Press
to save configuration.
9.
Go to Sensor Setup Menus → Sensor Setup E-Aux 1-8 - page 168 to
enable and configure set points, control, alarm, relays, and hold off time.
10. Press
to save configuration.

Data Logging

Menus → System →Data Logging
The 5X00 saves information in calibration, sensor, and event logs. Press
at the Run Screen to view the 25 most recent records for each log. The 5X00 sensor
log stores 2000 records. The calibration log can store 500 records. And the event
log stores 1250 records. Once a 5X00 log is full, it will begin overwriting the oldest
records with new records (data wrapping).
Configuring the 5X00
YSI 5X00
117