Hardware Installation

Correctly install 5X00 system components to ensure accurate data collection and
reliable operation. First, determine what monitor, control, and alarm functions
the 5X00 system will perform. Because system components are wired to the 5X00
instrument, it is important to determine where and how components will be installed
and configured. Preparing a system schematic (figure 3.1) is helpful.
Figure 3.1
Hardware installation steps (not all steps required for all applications):
Step 1 - install 5X00 instrument
Step 2 - locate and install sensor(s)
Step 3 - ground 5X00
Step 4 - wire power
Step 5 - wire sensors(s)
Step 6 - wire relays (output devices)
Step 7 - wire aux input devices (excluding temperature probes)
Step 8 - wire a network of 2 - 32 instruments
Step 9 - install and wire communication equipment
Step 10 - install and wire IOEM and REM instruments (optional). IOEM/REM
Installation information is included in a separate manual. See IOEM/REM
Installation and Safety Guide.
Installation and Wiring
YSI 5X00
26
See Chapter 4 Configuring 5X00 Systems - starting on 72 for directions on pro-
gramming 5X00 systems.
Tools and supplies needed for hardware installation:
Phillips-head screwdriver to remove 5X00 front panel to access I/O board
and lower board assembly
necessary tools for cutting mounting holes in mounting brackets (rail mount)
and/or control panel (panel mount)
5/32" Allen wrench (panel mount only)
electric drill and bits for mounting 5X00 (panel or rail) and for drilling rub-
ber grommets in bulkhead fittings
small flat-head screwdriver to terminate wires on I/O board pluggable
terminal strips
wrench to tighten bulkhead fittings following system component wiring
tweezers to replace the membrane gasket on the YSI 5422 sensor
wire strippers
distilled water for DO membrane installation
contactors may be required for relay wiring; see contactor wiring - page 55 .
Optional accessories for hardware installation:
surface mount hardware (included)
rail mount hardware
panel mount hardware
weather shield hardware
serial to ethernet device
wiring for network wiring
wiring not supplied with components and/or peripheral devices

Step 1 Install 5X00

Select a mounting location and method to mount the 5X00.

Location Considerations

The 5X00 must be located:
above any level where water damage can occur
away from extremely high or low temperature sources
away from vibrating surfaces
at least two 2 feet (~.7 m) from any high voltage conduit
Installation and Wiring