2.2 WARRANTY CARD ..........................2 2.3 BATTERIES..............................2 SECTION 3 PREPARING THE PROBE....................... 3 3.1 CHOOSING THE CORRECT PROBE...................... 3 3.1.1 YSI 5239 DO FIELD PROBE ......................3 3.1.2 YSI 5905 BOD PROBE ........................3 3.1.3 YSI 5739 DO PROBE ........................4 3.1.4 YSI 5718 DO FIELD PROBE ......................
Page 4
APPENDIX B REQUIRED NOTICE ......................30 APPENDIX C ACCESSORIES AND REPLACEMENT PARTS............31 APPENDIX D UNIT CONVERSION CHART ..................32 APPENDIX E OXYGEN SOLUBILITY TABLE..................33 APPENDIX F CALIBRATION VALUES TABLE .................. 35...
SECTION 1 INTRODUCTION The YSI 58 Dissolved Oxygen Meter is intended for field or laboratory use for dissolved oxygen and temperature measurement and water, but is also suitable for use in certain other fluids. See Section 4.4.2, Measuring Oxygen in Fluids Other than Water. The meter may be used with the YSI 5239 probe, 5905 probe or the 5700 Series probes.
When you unpack your new YSI 58 Dissolved Oxygen and Temperature System for the first time, compare the packing list with the contents of the shipping box. If there is anything missing or damaged, contact your YSI dealer from whom you purchased the YSI 58 or YSI Customer Service at 800-765-4974 or 937-767-7241.
YSI 5905 is a self-stirring BOD bottle probe. YSI 5739 is designed for field use and may be used with or without a stirrer. A 5740 probe cable or 5795A stirring cable is required for use with the 5739 probe. YSI 5795A has a dual-purpose cable to which both stirrer and probe are connected.
3.1.5 YSI 5750 BOD BOTTLE PROBE The YSI 5750 probe is similar in design to the 5739, but has a tapered body designed to fit standard US BOD bottles. It does not have a stirrer or cap membranes. Agitation of the sample must be provided by other means, such as a magnetic stirrer.
Four D-size carbon-zinc batteries that are housed in the lower battery compartment of the YSI 58 power the stirrer. When the stirrer will not be used for an extended period, these batteries should be removed from the instrument. PROBE INSTALLATION - YSI 5739 To disconnect the cable, unscrew the retaining nut and slide it down the cable to expose the connector.
5718 and 5739 probes. A selector switch inside the YSI 58 modifies the circuit for the membrane in use. This switch must be set to the position corresponding to the membrane in use. Facing the back of the instrument, with the back cover removed, the switch will be found on the top, right corner of the main PC board.
3. Hold the membrane cap and fill it at least half full with the electrolyte solution provided. Screw the membrane cap onto the probe moderately tight. A small amount of electrolyte should overflow. 4. Rinse off excess electrolyte from the probe with deionized water. 5. Reinstall the stir paddle. YSI Incorporated YSI 58 DO Meter...
3.3.3 MEMBRANE INSTALLATION FOR THE 5700-SERIES PROBES MEMBRANE INSTALLATION All YSI 5700 Series Probes have similar sensors and should be cared for in the same way. They are precision devices and require careful treatment if measurements of high accuracy are to be made.
Page 13
This ensures 100% humidity for accurate calibration and helps protect the probe against drying out in storage. The YSI 5905 and 5750 probes can be stored in a BOD bottle containing about 1” of water.
SECTION 4 OPERATION PRINCIPLES OF OPERATION The probes for use with the YSI 58 are a Clark type voltammetric (polarographic) sensor of dissolved oxygen. An oxygen permeable membrane covers an electrolytic cell consisting of a cathode and an anode. This membrane acts as a diffusion barrier and an isolation barrier preventing the fouling of the cathode surface by impurities in the environment.
8 hours per day operation. TURNING THE INSTRUMENT ON The YSI 58 may be used in a vertical, horizontal or tilted position. It may be carried or moved during use without affecting its accuracy or stability of measurement.
4.3.1 AIR CALIBRATION Air Calibration is the quickest and by far the simplest calibration technique. Experience has shown that it is reliable and is recommended for the YSI 58. Air Calibrate the YSI 58, with any field probe as follows: Set the function switch to % Mode.
Readjust zero if necessary. Check that the SALINITY knob is set at zero. Adjust the 02 CALIB control to the calibration value determined in the foregoing step. Wait two minutes to verify stability; readjust if necessary. YSI Incorporated YSI 58 DO Meter...
The YSI 5795A Submersible Stirrer supplies the necessary stirring for the 5739 probe. The YSI 5905 BOD probe has its own line powered stirrer for laboratory use. When the 5750 BOD probe is used, stirring must be provided such as with a magnetic stirrer.
4.4.2 MEASURING OXYGEN IN FLUIDS OTHER THAN WATER The YSI 58 is normally used for measuring the oxygen content of naturally occurring waters and wastewaters. The % air saturation feature of the instrument additionally permits 02 measurement in some non-water fluids including air, most gases, foods and some non-aqueous liquids.
Output at full scale is 1VDC. (0.0005V = 1 digit). Use a recorder with input impedance of at least 50K OHM, and operate it with the terminals ungrounded. A recorder output plug is provided. Cable for connecting the recorder must be supplied by the user. YSI Incorporated YSI 58 DO Meter...
2. Erratic readings are a result of loose, wrinkled, damaged, or fouled membranes, or from large (more than 1/4 of the circumference of the probe) bubbles in the electrolyte reservoir. If erratic YSI Incorporated YSI 58 DO Meter...
3 times or until all silver deposits are removed and the gold appears to have a matte finish. If the cathode remains tarnished, return the probe for service. Temperature Sensor Anode (Silver) Cathode (Gold) Figure 8 SILVER ANODE YSI Incorporated YSI 58 DO Meter...
If the silver anode becomes contaminated, successful calibration can be prevented. Soak the probe overnight in a 3% ammonia solution; rinse with deionized water, and follow the steps to install a new membrane. If it is not possible to calibrate, return the probe to an authorized YSI Repair Center. PROBE STORAGE For long term storage (4 weeks), remove the membrane, thoroughly rinse the sensor with deionized or distilled water and install a new membrane.
If the probe is calibrated in water-saturated air, then type 3B errors (humidity), the largest error of all types, is virtually eliminated and the maximum possible error is in the order of 0.1 mg/L for the case of calibrating around 25°C. YSI Incorporated YSI 58 DO Meter...
Change membrane Clean probe (see instructions under Electrode Cleaning) Unstable readings Allow sufficient time for temperature and oxygen stabilization Allow 60 sec for reading to stabilize Change membrane Clean probe (see instructions under Electrode Cleaning) YSI Incorporated YSI 58 DO Meter...
Page 26
Check if stir paddle is pushed all 3.1.2 the way in Replace motor Bent stir paddle Replace stir paddle Membrane cap is tight on the Use pliers to loosen the cap (turn 5905 counter clockwise while facing sensor tip) YSI Incorporated YSI 58 DO Meter...
SECTION 8 WARRANTY AND SERVICE INFORMATION YSI 58 Dissolved Oxygen Meters are warranted for two years from date of purchase by the end user against defects in materials and workmanship. YSI 58 probes and cables are warranted for one year from date of purchase by the end user against defects in material and workmanship.
Page 28
AUTHORIZED U.S. SERVICE CENTERS For information on the nearest authorized service center contact: YSI Technical Support • 1725 Brannum Lane • Yellow Springs, Ohio • 45387 Phone: (800) 765-4974 (US) • +1 (937) 767-7241 • Email: environmental@ysi.com www.ysi.com YSI Incorporated...
Page 30
We will contact the sender for instructions as to the disposition of the equipment. Disposition costs will be the responsibility of the sender. When service is required, either at the user's facility or at YSI, the following steps must be taken to insure the safety of our service personnel.
Page 31
Model No. of Device ______ Lot Number _________ Contaminant (if known) _____________________ Cleaning Agent(s) used _____________________ Radioactive Decontamination Certified? (Answer only if there has been radioactive exposure) ___ Yes ___ No Cleaning Certified by _________________________ Name Date YSI Incorporated YSI 58 DO Meter...
+ 0.5% of reading, + 2 least significant digits (in the 0.01 mg/L mode) Recorder Output: Voltage: 0 to 1 volt, full scale Accuracy: + 0.4 % of full scale, + 1 least significant digit (in the 0.01 mg/L mode) Minimum Load Impedance: 50 K OHM YSI Incorporated YSI 58 DO Meter...
Page 33
10 seconds at a low constant temperature of 30°C with the 5775 membrane. DO response and low temperature and low DO is typically 90% in 30 seconds. YSI 5776 High Sensitivity membranes may be used to improve response at low temperature and DO concentrations.
The user may find the following booklet, prepared by the Federal Communications Commission, helpful: "How to Identify and Resolve Radio-TV Interference Problems." This booklet is available from the U.S. Government Printing Office, Washington, DC 20402, Stock No. 0004-000-00345-4. YSI Incorporated YSI 58 DO Meter...
APPENDIX C ACCESSORIES AND REPLACEMENT PARTS The following parts and accessories are available from YSI or any Franchise Dealer authorized by YSI. YSI Order Number Description 5905 Self-Stirring BOD Bottle Probe 5750 Non-Stirring BOD Bottle Probe 5740 –10 Detachable leads for use with the 5739 probe only. Comes in 10, 25, 50, 100, 150, 200 feet lengths.
Feet Meters Multiply by 0.3048 Meters Feet Multiply by 3.2808399 Degrees Celsius Degrees Fahrenheit (9/5 C)+32 Degrees Fahrenheit Degrees Celsius 5/9 ( F-32) Milligrams per liter (mg/l) Parts per million (ppm) Multiply by 1 YSI Incorporated YSI 58 DO Meter...