Summary of Contents for YSI 550A

  • Page 1 Operations 550A Manual Dissolved Oxygen Instrument...
  • Page 3: Table Of Contents

    CONTENTS CONTENTS............................iii GENERAL DESCRIPTION........................1 WARRANTY ............................2 FEATURES OF THE YSI 550A ......................3 PRINCIPLES OF OPERATION......................5 PREPARING THE PROBE ........................5 DISSOLVED OXYGEN CALIBRATION ..................6 SALINITY COMPENSATION CALIBRATION ................7 PROBE OPERATION ...........................8 PROBE ELECTRODE MAINTENANCE..................9 ACCESSORIES/REPLACEMENT PARTS ..................10 SPECIFICATIONS ..........................11 TROUBLESHOOTING........................12 CONTACT INFORMATION ......................13 REQUIRED NOTICE..........................14...
  • Page 5: General Description

    The YSI 550A DO Instrument is designed for field use and is available with cable lengths of 12, 25, 50, or 100 feet (3.5, 7.5, 15, 30.5 meters). The body of the probe has been manufactured with stainless steel to add rugged durability and sinking weight.
  • Page 6: Warranty

    Your purchase of this quality instrument will then be recorded in YSI’s customer database. Once your purchase is recorded, you will receive prompt, efficient service in the event any part of your YSI 550A DO Instrument should ever need repair.
  • Page 7: Features Of The Ysi 550A

    % measurement mode. BATTERIES The YSI 550A DO Instrument is powered by 4 C-size alkaline batteries. A new set of alkaline batteries will provide approximately 2000 hours of continuous operation. When batteries need to be replaced, the LCD will display a "LO BAT"...
  • Page 8 This environment is ideal for dissolved oxygen calibration and for storage of the probe during transport and non-use. The YSI 550A DO Instrument’s storage chamber can be conveniently used from either side of the instrument.
  • Page 9: Principles Of Operation

    PREPARING THE PROBE MEMBRANE CAP INSTALLATION The YSI 550A DO probe is shipped with a dry, protective membrane. Before using the instrument for the first time, remove the protective cap and replace it with a new one following these instructions: Remove the probe sensor guard to access the probe tip.
  • Page 10: Dissolved Oxygen Calibration

    DISSOLVED OXYGEN CALIBRATION Dissolved oxygen calibration must be done in an environment with known oxygen content. The YSI 550A DO Instrument can be calibrated in either mg/L or % saturation. Sections below include instructions on how to calibrate in either mode.
  • Page 11: Salinity Compensation Calibration

    2. Use the UP ARROW and DOWN ARROW keys to adjust the salinity value to that of the samples you will be measuring, 0-70 ppt. Press the ENTER key to save the calibration. Press Mode to return to dissolved oxygen measurement Environmental Y S I YSI 550A Page 7...
  • Page 12: Probe Operation

    PROBE OPERATION NOTE: The YSI 550A DO Instrument should not be used in a purpose other than that specified by YSI Incorporated. See Warranty for details. STIRRING It is important to recognize that a very small amount of oxygen dissolved in the sample is consumed during probe operation.
  • Page 13: Probe Electrode Maintenance

    For correct probe operation, the gold cathode must be textured properly. It can become tarnished or plated with silver after extended use. The gold cathode can be cleaned by using the YSI 5238 Probe Reconditioning Kit or 400 grit wet/dry sandpaper. Never use chemicals or abrasives not recommended or supplied by YSI.
  • Page 14: Accessories/Replacement Parts

    DO sensor module. Also, if the hex screw was removed completely, make sure that it is not cross-threaded when replaced. The YSI 559 DO module is shipped with a dry, protective membrane. Before using for the first time, remove the protective cap and replace it with a new one following the instructions for Membrane Cap Installation.
  • Page 15: Specifications

    12, 25, 50, and 100-foot lengths (3.5, 7.5, 15, 30.5 meter lengths) Waterproof to IP-67 Other Features: High-impact resistance Push-button calibration Built-in calibration chamber Large back-lit display Low battery indicator on display Manual salinity input CE-compliance Environmental Y S I YSI 550A Page 11...
  • Page 16: Troubleshooting

    D. Clean probe electrodes. E. Return system for service. 6. Main display reads “Over” or A. Replace membrane and electrolyte. Recalibrate. “Undr” during calibration. B. Clean probe electrodes. C. Return system for service. Environmental Y S I Page 12 YSI 550A...
  • Page 17: Contact Information

    “EEP” CONTACT INFORMATION YSI offers a wide range of customer assistance and technical support functions to ensure that you have the information required to use our products. Contact YSI if you need assistance or have questions regarding any YSI Product. Business hours are Monday through Friday, 8AM to 5PM ET.
  • Page 18: Required Notice

    Note: While testing to EN-61000-4-6, Conducted RF Immunity, per Table A.1 of EN61326, Electrical Equipment for Measurement, Control and Laboratory Use, the YSI 550A exhibited an ERROR 8 message from 8.6 MHz 22.8 MHz at induced RF voltages of 3-Volts to 1-Volt RMS on the 25-foot probe cable. If you observe this interference please relocate the probe-cable away from heavy industrial equipment power and control cables or communications equipment cables which may be causing the interference.