Configure E-Aux; Probe Name - YSI MultiDO 5400 User Manual

Monitor, control, & alarm instruments
Table of Contents

Advertisement

2.
Enter Aux menu
Menus → System → Configure Aux →Aux 3-6
3.
Configure temperature input type and name.
▲ ▼
Use
4.
to scroll and highlight submenus.
Press
5.
to select.
6.
Press
to save settings and return to Run Screen.
7.
Go to Sensor Setup Menus → Sensor Setup Temperature for Aux 3-6
and ODO - 168 to configure temperature system.
8.
Press
to save settings and return to Run Screen.
Note:
-
The Sensor Setup menu - page 150 is automatically enabled when an aux
channel is configured for temperature.

Configure E-AUX

Menus → System → Configure E-Aux 1-8
NEW
All 5X00 instruments have eight E-Aux systems. E-Aux systems are sometimes
referred to as virtual system because unlike Aux systems which use six onboard
(Aux) inputs for wiring input devices, E-Aux systems use inputs that are wired to a
networked IOEM. 5X00 E-Aux systems mimic aux systems but use a remote IOEM
wiring location for the actual input. E-Aux inputs 1-8 can be configured for digital,
analog ( 0-1V,DC 0-5VDC, or 4-20 mA) or temperature inputs. System → Configure
E-Aux 1-8 menu selections change based on input type selection. Configure suffix,
minimum (Min), maximum (Max), Input Scale Min, Input Scale Max and display
format menus for non temperature analog inputs. To configure setpoints,control,
alarms and relays see specific sensor setup menus for configured input type For E-
Aux Analog Sensor Setup - page 166. For E-Aux Temperature Sensor Setup - page
168. For E-Aux Digital Sensor Setup - page 165.
Configuring the 5X00
YSI 5X00
108
Notes:
-
Configure E-Aux systems to monitor, control and/or alarm IOEM inputs the
same way 5X00 aux sensor systems function.
-
E-Aux systems have data and event logging functionality.
-
E-Aux system datalogs are not visible at the instrument.
-
At the Run Screen "Offline" display at an E-Aux system means the config-
ured IOEM input value that the associated 5X00 instrument E-Aux system is
configured for is not recognized. An example - if an IOEM channel mode is
configured to digital and the input signal is a 0-1V signal, the reading at the
run screen will be "Offline";

Probe Name

Menus → System → Configure E-Aux →E-Aux 1-8→Probe Name
Use the alpha/numeric keypad to name E-Aux inputs. Configured name(s) appear
throughout menus and at Run Screen - figure 4.7- page 106.
Note:
-
E-Aux names can also be configured at Menus → System → Name Devices/
Sensors>Name E-Aux>E-Aux 1-8.
Configuring the 5X00
YSI 5X00
109