Detaching The Collimator Mount - Brainlab RT/RS Hardware User's Manual

Stereotactic hardware
Table of Contents

Advertisement

HARDWARE OVERVIEW
2.3.3

Detaching the Collimator Mount

Background
Never change the collimator mount while a person is beneath the collimator mount.
Detaching the Collimator Mount (Siemens/Varian)
Steps
Further Information
Please refer to the Linac manufacturer's manual for further information on attaching/ detaching
accessories. The base plate of the collimator mount as well as the accessory locking mechanism
varies depending on the manufacturer of the Linac.
Hardware User Guide Rev. 4.9 RT / RS Stereotactic Hardware
1.
Position the gantry at 0°/180° (according to the IEC 61217 scale).
2.
Remove the conical collimator (see Page 43).
Turn the Linac collimator so that the opening of the accessory tray slot points towards
3.
you.
Unscrew the two fixation screws (③ on Figure 11).
4.
5.
Make sure that the accessory locking mechanism of the Linac is released.
6.
Slide the collimator mount completely out of the accessory tray slot and remove it.