CLEANING, DISINFECTION, AND STERILIZATION
7.2.6

Stereotactic Target Positioner, Target Positioner for Leksell Headring

Maintenance and Storage
The target positioner carrier plates are not scratch proof and must be handled with care.
Do not store or use the target positioner carrier plates in humid places, under conditions of
constant humidity, or above 80°C (176°F).
Cleaning and Disinfection
It is recommended that the target positioner is periodically cleaned using a soft and dry
cloth. Stubborn stains may be removed with a slightly moistened cloth.
Sterilization
The target positioner carrier plates must not be sterilized under any circumstances. Do not
use cleaning solutions, solvents or ultrasonic cleaners to clean the plates.
Hardware User Guide Rev. 4.9 RT / RS Stereotactic Hardware