CLEANING, DISINFECTION, AND STERILIZATION
7.2

Stereotactic Hardware

7.2.1

Overview

General Information
This information applies to all RT/RS Stereotactic Hardware. Instructions for specific hardware is
provided in subsequent sections.
Handling and Storage
Store RT/RS Stereotactic Hardware in a clean and dry area, safe from infectious products,
impacts, scratches, and falling objects.
Maintenance
RT/RS Stereotactic Hardware does not require particular maintenance. Maintain its cleanliness by
using a smooth duster and the cleaning products used in the radiotherapy department.
Cleaning and Disinfection
The following are required:
• A clean and smooth duster
• A pair of gloves
• A diluted disinfectant detergent, according to the rules in the radiotherapy department (or
according to the rules of the manufacturer)
The cleaning and disinfectant products must comply with the following:
• They must not be too acidic or too basic (pH-value between 6 and 8)
• They must not contain caustic or corrosive substances
• Their composition must not change the plastic structure
• They must not be too abrasive
Never use abrasive products (scourers or pad scourers) or solvents.