To change the serial port settings for printing:
1. Set the rotary dial on the Main Unit to RESULTS.
2. Press the up- and down-arrow keys until your
report selection is highlighted. See Section 7.2 for
specific instructions.
3. Press the F4 button.
4. Press the F1 button to increase the baud rate from
1200 bps to up to 38,400 bps.
5. Press the F2 button to select the flow control:
• XON/XOFF (software)
• HARDWARE (CTS/DTR)
• NONE (for no flow control)
6. Press the F3 button to select the output format as
follows:
• TEXT for a pre-formatted report ready for
printing.
• COMMA for comma-separated values (CSV) to
import data to a database or spreadsheet.
CHAPTER 7: Test Results