FIBER METER
• Switching the Fiber Meter handsets from Remote
to Main operation.
The Fiber Meter compensates for changes in ambient
temperature. For optimum performance, allow the test
system to adjust to the ambient temperature of the test
area before setting the Autotest reference.
You'll also need to re-establish the reference when there
has been a significant change in temperature since the
reference was last established. If the Fiber Meter is used
over long intervals of time, you should re-establish the
reference every couple of hours to allow for changes in
ambient temperature.
Use test jumpers that are of the same type of fiber as the
cable you will test (either 50/125 µm or 62.5/125 µm).
You can use either of the following two reference setting
methods:
• Method B per TIA/EIA-526-14A (OFSTP-14A);
this is described in Section 5.2.1.
• Method A per TIA/EIA-526-14A (OFSTP-14A);
this is described in Section 5.2.2.
42