5.2.2
ExacTrac Startup Using the Timer Switch
General
The timer switch enables you to power up the system before use. The camera requires a warm-up
Information
time of 60 minutes before the system can be calibrated or patients treated.
Timer Switch
Programming the
The timer switch can be programmed with a fixed startup date and time, or with startup information
Timer Switch
for several days. Detailed operating instructions are provided in the corresponding manufacturer in-
struction manual provided with your system.
48
If your system is fitted with a timer switch, Brainlab strongly recommends using this
switch to power up the system at least 60 minutes before use.
Failure to observe a warm-up time of 60 minutes following system start will result in
tracking inaccuracies greater than 2 mm.
Figure 19
Clinical User Guide Rev. 1.1, ExacTrac Infrared Monitoring Ver. 1.0