Documentation Of Tests - Brainlab EXACTRAC INFRARED MONITORING Clinical User Manual

Table of Contents

Advertisement

8.3.2

Documentation of Tests

General
In accordance with the standard DIN VDE 0751-1, Brainlab recommends logging all measurements
Information
for the purposes of comparison. In the event of significant deviations compared with previous mea-
surements, system failure is likely to occur. In such cases, further detailed investigation should be
completed in cooperation with Brainlab support.
Required Records
The test protocol must contain at least the following information:
• Description of executing organization
• Name of technician carrying out test
• Serial number and exact system description (version)
• Test description and results of:
• Overall evaluation
• Date and signature
File all printouts and note the highest (most dangerous) electrical impedance in the test protocol.
Clinical User Guide Rev. 1.1, ExacTrac Infrared Monitoring Ver. 1.0
- Visual test
- Protective earth resistance test
- Leakage current test
- Functionality test
85