Connection
Restrictions
NOTE: Unless otherwise specified, the use of multiple socket outlet(s) is not permitted.
Notes on
Electromagnetic
Compatibility
Should electromagnetic interference occur, the operator must take the following measures:
• Reposition or relocate the affected device(s)
• Increase the distance between the equipment and the affected device
• Attach the affected device to a different power source
• Contact Brainlab support for further advice
28
Do not connect miscellaneous equipment to the ExacTrac data interfaces or the inte-
grated power rail. Only equipment specified by Brainlab may be connected to the sys-
tem.
Do not connect the ExacTrac system components directly to the mains power supply.
To guarantee electrical safety for both the user and patient, ExacTrac system compo-
nents may only be powered via the integrated power rail.
Additional equipment connected to medical electrical equipment must comply with
the respective IEC or ISO standards (e.g., IEC 60950 for data processing equipment
and IEC 60601-1 for medical equipment). Furthermore, all configurations shall comply
with the requirements for medical electrical systems (see IEC 60601-1-1 or clause 16
of the 3rd edition of IEC 60601-1, respectively). Anyone connecting additional equip-
ment to medical electrical equipment configures a medical system and is therefore re-
sponsible that the system complies with the requirements for medical electrical
systems. Please note that local laws take priority over the above mentioned require-
ments. In doubt, consult your local representative or the technical service department.
ExacTrac is certified according to IEC 60601-1-2 (EN55022, Class A). Although this
system satisfies the required emission limits for medical devices, it may induce elec-
tromagnetic interference in medical and non-medical devices or disrupt radio commu-
nication.
Do not operate ExacTrac if electromagnetic interference is suspected in the treatment
room.
Brainlab is not responsible for interference caused by the use of interconnect cables
other than those recommended.
Clinical User Guide Rev. 1.1, ExacTrac Infrared Monitoring Ver. 1.0