8.3.3

Visual Inspection

General
Before performing further inspections, a qualitative validation of the connections and wirings of all
Information
potential equalization conductors within the system should be conducted for safety reasons.
Components to be
• All connections should be verified in order to ensure that they have not become loose.
Reviewed
• All potential equalization wirings should be checked in order to ensure that they are not worn
86
out or damaged, in particular where those wirings are exposed to mechanical stress, e.g. within
the ceiling mounted monitor arm.