Memory Clearing, Sanitization and Removal Procedures

Instrument Sanitization Procedures

Instrument Sanitization Procedures
This section includes flowcharts that describe how to sanitize an instrument by
physical removal and replacement of the Disk Drive.

Application License Key Storage

Note that License keys for all Applications are stored in EEPROM on the A7
Midplane Assembly (as described in Item 8 of
when replacing the Disk Drive, you do not need to back up and restore the
license keys.

Replacement of Disk Drive

For details of how to archive or restore the instrument's calibration files (Steps
3, 12 and 16 in the flowchart), see
Data Files" on page
For details of how to remove the Disk Drive (Step 6), see
Procedure" on page
IMPORTANT
Ensure that the instrument software revision on the secondary non-classified
disk drive matches that of the classified drive in the controlled area.
If the non–classified drive receives a software upgrade, or if the non-classified
drive is replaced with a drive that contains a newer revision of software than
that of the classified drive, the classified drive will require a software upgrade
(inside the controlled area) to match the non-classified drive.

Items Required

To perform these tasks in a controlled area, you will need:
— 1x Instrument
— 2x Disk Drives for the instrument
— 1x One-time writable media, plus 1x compatible writing device for the

Procedure

Steps below that appear on a yellow background take place in the controlled
area. All other steps take place in a non-controlled area.
Step
1
Security Features & Document of Volatility
23.
25.
non-controlled area and 1x reading device for the controlled area. See
"Selection of USB Memory Device" on page 23
Task
Receive new or recalibrated instrument from factory/service center.
Table 4-1 on page
"Archiving and Restoring Factory Calibration
for details.
12). Therefore,
"Disk Drive Removal
21