function.
Press SourceA, and rotate the multifunction knob
Source A. Then press down the knob to select the channel. You can also press
SourceA continuously or enable the touch screen to select the channel. The
available channels for Source A and Source B are CH1-CH2 and D0-D15.
Note:
Only the channels that have been enabled currently can be selected.
If Source A (or Source B) selects CH1, press More → Thre.CH1 to set the
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in "Parameter Setting Method".
If Source A (or Source B) selects CH2, press More → Thre.CH2 to set the
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in "Parameter Setting Method".
Press Offset, and then rotate the multifunction knob
offset of the operation results. You can also enable the touch screen and use the
"Drag" gesture to adjust the vertical offset.
Press Size, and then rotate the multifunction knob
size. Press down the knob to select it. You can also press Size continuously or
enable the touch screen to select it. The available sizes include "Small",
"Medium", and "Large".
Press AutoSetting to adjust the vertical scale and offset of the operation
results to the optimal value according to the current configuration, so as to
better observe.
Press More → Sensitivity to set the sensitivity of the digital signal converted
from the analog signal on the source. For setting methods, refer to descriptions
in "Parameter Setting Method".
Press More → Label to set the label for math operation results. For details,
refer to descriptions in "Math Operation Label".

Intg

Calculates the integral of the selected source. For example, you can use integral to
measure the area under a waveform or the pulse energy.
Press Math → Math1 → Operator to select "Intg":
Press Operation continuously to enable or disable the "Intg" operation
function.
Press SourceA, and rotate the multifunction knob
Source A. Then press down the knob to select the channel. You can also press
SourceA continuously or enable the touch screen to select the channel. The
available channels include CH1-CH2 and Ref1-Ref10.