desired baud rate.
Set the signal type
Press Signal and rotate the multifunction knob
that matches the actual FlexRay bus signal. Press down the knob to select it.
You can also press Signal continuously or enable the touch screen to select
it. The available signal types include BP, BM, and RX/TX.
Sample Position:
Sample position is a point within a bit's time. The oscilloscope samples the
bit level at this point. The sample position is represented by the proportion
of "the time from the start of the bit to the sample position" to the "bit time",
as shown in the figure below.
Press Sample Position, then rotate the multifunction knob
pop-up numeric keypad to set the sample position. The settable range is
from 10% to 90%.
5. Display-related setting
Press Display to enter the display setting menu.
Set the display format
Press Format, rotate the multifunction knob
format of the bus data. Then, press down the knob to select it. You can also
press Format continuously or enable the touch screen to select it. The
available display formats of the bus data are "Hex", "Dec", "Bin", and
"ASCII".
Adjust the vertical position of the bus
Press Position, and then rotate the multifunction knob
vertical display position of the bus.
Set the label display
Press Label to enable or disable the label display of the FlexRay decoding
bus. When enabled, the bus label "FlexRay" will be displayed at the
upper-left side of the bus (when the bus display is enabled).
11-34
One Bit
60%
Figure 11-12 Sample Position
Chapter 11 Protocol Decoding
to select a signal type
70%
80%
to select the display
or use the
to adjust the
MSO5000-E User Guide