Sqrt - Rigol MSO5000-E Series User Manual

Digital oscilloscope
Hide thumbs Also See for MSO5000-E Series:
Table of Contents

Advertisement

RIGOL
Press Offset to set the vertical offset of the operation results. For setting
methods, refer to the descriptions in "Parameter Setting Method". You can
also enable the touch screen and use the "Drag" gesture to adjust the vertical
offset.
Press Scale to set the vertical scale of the operation results. For setting methods,
refer to the descriptions in "Parameter Setting Method". You can also enable
the touch screen and use the "Pinch & Stretch" gesture to adjust the vertical
scale.
Press AutoSetting to adjust the vertical scale and offset of the operation
results to an optimal value based on the current configuration, so as to better
observe.
Press More → Smooth to set the number of smoothing times in differential
operation. For setting methods, refer to descriptions in "Parameter Setting
Method".
Note:
As the high-frequency component of waveform data has a great impact
on the differential operation, you need to adjust the parameter "Smooth" to a
larger value, so as to restore the low-frequency component.
Press More → Invert to enable or disable the inverted display function of the
waveform.
Press More → Label to set the label for math operation results. For details,
refer to descriptions in "Math Operation Label".
Press More → Color Grade to enable or disable the color grade display of the
operation results. When enabled, different colors are displayed on the screen to
indicate the times of data acquisition or acquisition probability.
Press More → Reset ColorGrade to clear the color grade display and display
the color grade again.
Tip
Because differentiation is very sensitive to noise, it is helpful to set acquisition
mode to "Average".

Sqrt

Calculates the square root of the selected source point by point and displays the
results.
Press Math → Math1 → Operator to select "Sqrt":
Press Operation continuously to enable or disable the "Sqrt" operation
function.
Press SourceA, and rotate the multifunction knob
Source A. Then press down the knob to select the channel. You can also press
SourceA continuously or enable the touch screen to select the channel. The
6-16
Chapter 6 Operations and Measurements
to select the channel for
MSO5000-E User Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents