Chapter 6 Operations and Measurements
or enable the touch screen to select the channel. The available channels for
Source A and Source B are CH1-CH2 and D0-D15.
Note:
Only the channels that have been enabled currently can be selected.
If Source A (or Source B) selects CH1, press More → Thre.CH1 to set the
➢
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in "Parameter Setting Method".
If Source A (or Source B) selects CH2, press More → Thre.CH2 to set the
➢
threshold of Source A (or Source B) in logic operation. For setting methods,
refer to descriptions in "Parameter Setting Method".
Press Offset, and then rotate the multifunction knob
⚫
offset of the operation results. You can also enable the touch screen and use the
"Drag" gesture to adjust the vertical offset.
Press Size, and then rotate the multifunction knob
⚫
size. Press down the knob to select it. You can also press Size continuously or
enable the touch screen to select it. The available sizes include "Small",
"Medium", and "Large".
Press More → AutoSetting to adjust the vertical scale and offset of the
⚫
operation results to the optimal value according to the current configuration, so
as to better observe.
Press More → Sensitivity to set the sensitivity of the digital signal converted
⚫
from the analog signal on the source. For setting methods, refer to descriptions
in "Parameter Setting Method".
Press More → Label to set the label for math operation results. For details,
⚫
refer to descriptions in "Math Operation Label".
"NOT" Operation
Performs logic "NOT" operation on the waveform voltage values of the specified
sources point by point and displays the results. In operation, when the voltage value
of the source channel is greater than the threshold of the corresponding channel, it is
regarded as logic "1"; otherwise it is logic "0". The results of logic "NOT" operation of
two binary bits are as shown in Table 6-5.
Table 6-5 Logic "NOT" Operation
A
!A
0
1
1
0
Press Math → Math1 → Operator to select "!A":
Press Operation continuously to enable or disable the "NOT" operation
⚫
MSO5000-E User Guide
RIGOL
to adjust the vertical
to select the waveform
6-13